Researchers Find Cross-System Activation Possible Not Recommended

Researchers have found active bone conduction implant processors from different systems can technically activate one another, but the practice is not recommended due to significant safety concerns.

Researchers have found active bone conduction implant processors from different systems can technically activate one another, but the practice is not recommended due to significant safety concerns.

A study published in November 2026 investigated the compatibility of the Bonebridge (MED-EL) and Sentio (Oticon Medical) systems.

The investigation, titled “Can active bone conduction implant processors of two different systems be interchanged? A technical and clinical feasibility study”, was published in the journal Zeitschrift für Medizinische Physik.

The authors were Torsten Rahne of the University Medicine Halle, Department of Otorhinolaryngology, Head and Neck Surgery, Dawid Brüning of the same department, Laura Fröhlich of the University Medical Center Bonn, Department of Otorhinolaryngology, and Oliver Dziemba of the Department of Otorhinolaryngology, Head and Neck Surgery, University Medicine Greifswald.

The study evaluated the technical feasibility and preliminary clinical applicability of cross-system processor-implant combinations, specifically using the Bonebridge BCI 602 and Sentio Ti implants. A Bonebridge BCI 602 and a Sentio Ti implant were coupled to a skull simulator and implant phantoms for testing.

The study concluded that while cross-system activation was technically possible across all tested combinations, the resulting performance varied substantially. Skull simulator measurements revealed force output differences of up to 21.6 decibels between native and non-native pairings, indicating a lack of electro-mechanical optimization.

“Nevertheless, substantial differences in output characteristics and the lack of systematic electrical, thermal, and long-term safety evaluations mean that clinical interchangeability cannot be recommended,” the study’s authors stated. “Further technical and clinical evaluation is required before cross-system use can be considered.”

In the two clinical assessments, participants achieved successful fitting and speech understanding with the non-native processors. Their aided thresholds and hearing outcomes were generally comparable between native and non-native configurations, though both participants showed a preference for the Sentio 1 Mini processor in some tests.

A separate study examined performance of a different piezoelectric bone conduction device, the Osia OSI300. This research focused on implant positioning, finding that location significantly affects cochlear promontory velocity and transcranial attenuation, particularly at higher frequencies.