Skull simulator measurements revealed 4PTA force-output differences reaching up to 21.6 dB between native and non-native bone conduction processors, according to a study published in November 2026 in the Zeitschrift für Medizinische Physik.
Active transcutaneous bone conduction implant processors can physically activate devices from different manufacturers, but distinct output differences mean they cannot be clinically interchanged.
Researchers Torsten Rahne, Dawid Brüning, Laura Fröhlich, and Oliver Dziemba evaluated cross-system processor-implant combinations using equipment from MED-EL and Oticon Medical. Active transcutaneous bone conduction systems are built as closed hardware units where external audio processors work solely with matching implanted transducers, prompting the team to test whether components could mix.
Feedback-related interference also affected measurements at high input levels, according to the findings from University Medicine Halle, the University Medical Center Bonn, and University Medicine Greifswald.
Phantom measurements confirmed audibility across all tested combinations despite the output discrepancies. Two implanted patients participated in acute experimental fittings using the alternative processor systems, achieving successful fitting and open-set speech understanding.
Aided thresholds, speech recognition performance, and subjective hearing outcomes remained generally comparable across both native and non-native setups. Both participants indicated a preference for the Sentio 1 Mini processor in at least some outcome measures.
Torsten Rahne and Dawid Brüning are affiliated with the Department of Otorhinolaryngology, Head and Neck Surgery at University Medicine Halle in Halle (Saale), Germany. Laura Fröhlich works at the Department of Otorhinolaryngology at the University Medical Center Bonn in Bonn, Germany.
Oliver Dziemba is based at the Department of Otorhinolaryngology, Head and Neck Surgery at University Medicine Greifswald in Greifswald, Germany. The study authors said that while short-term speech perception and technical activation are possible, missing systematic electrical, thermal, and long-term safety evaluations prevent any recommendation for clinical interchangeability.
Ipsilateral velocity at the manufacturer-specified location immediately posterior to the pinna was 6.84 and 6.66 dB higher than alternative posterior positions in intact mastoid conditions. Transcranial attenuation values of 12 to 22 dB at 3,000 and 6,000 Hz at that primary location demonstrated strong ipsilateral vibration. Smaller attenuation values were recorded at the alternative testing positions.
Velocity shifts following mastoidectomy remained within 3.7 dB across frequencies from 250 to 1,500 Hz, though measurements proved inconsistent at higher frequencies.