Study Finds Lab Plastic Additives Contaminate Samples, Almería Researchers Say

A study found that laboratory plastic additives contaminate samples at levels that can skew results.

GC and LC–HRMS Track Laboratory Plastic Additive Leaching

A study found that laboratory plastic additives contaminate samples at levels that can skew results.

Researchers at the University of Almería in Spain conducted what they believe is the first comprehensive assessment of non-phthalate plastic additive (NPPA) contamination from common laboratory supplies. The study, published in Analytica Chimica Acta, examined 28 different additives across various consumables and extraction workflows, finding that some materials introduce significant background noise into sensitive analyses.

“The lack of quantitative information on NPPA contamination introduced by routine laboratory consumables hampers the reliability of trace-level plastic additive analysis,” the researchers said.

The investigation focused on “blank” tests run with centrifuge tubes, pipette tips, syringes, syringe filters, and extraction salts, using ethyl acetate and acetonitrile as solvents. Analysis was performed with both gas chromatography and liquid chromatography, each paired with high-resolution mass spectrometry.

Study Finds Lab Plastic Additives Contaminate Samples, Almería Researchers Say
Photo: sciencedirect.com

The study identified syringe-filter assemblies as the main source of contamination. Nylon membrane filters were found to release up to 218.7 nanograms (ng) of total NPPAs into ethyl acetate extracts. Polyetetrafluoroethylene (PTFE) filters released approximately 70.1 ng, according to the findings published in ScienceDirect.

This stark difference highlights material choice as a critical factor in minimizing background contamination. The results showed that for most materials tested, samples prepared with ethyl acetate picked up more contamination than those prepared with acetonitrile.

Component Total NPPA Leached (ng)
Nylon syringe filter 218.7
PTFE syringe filter 70.1
QuEChERS salt kit (HCPK) ~1840x higher than individually weighed salts

Storage conditions also proved influential. The researchers said that technical-grade acetone stored in plastic containers carried about 13 times more background contamination than LC-MS-grade acetone kept in glass.

Another major discovery was the contamination level in commercially pre-packaged QuEChERS salt kits. They were substantially dirtier than salts weighed out individually by researchers. The chemical 1-hydroxycyclohexyl phenyl ketone (HCPK), used to help plastics harden, showed up at levels approximately 1840 times higher in the pre-packaged kits compared to hand-weighed salts.

The authors of the study in Analytica Chimica Acta reported that the high contamination from these kits was a key factor in their conclusion. They found that consumable selection can have a greater impact on background levels than the extraction workflow itself.

The study stressed that practical mitigation strategies are important for improving reliability. These include careful material selection and blank correction for trace-level plastic additive determinations in research.

The researchers assessed the cumulative impact of material selection across entire sample-preparation workflows. They found that best-case and worst-case contamination scenarios differed by nearly one order of magnitude in gas chromatography-based workflows. For QuEChERS-based liquid chromatography workflows, the difference was more than two orders of magnitude.

Raquel Capilla-Flores and her team from the University of Almería explained that this demonstrates the strong influence of material choices on cumulative contamination. They said that their work provides the first quantitative assessment of NPPA contamination from routine laboratory consumables across complete sample-preparation workflows.